MedSystem III Pump
Alarm: Check Fluid Side
Meaning:
- Possible upstream restrictions to flow
Response:
- Check tubing between container and pump for closed regulating clamp, closed vent (with unvented container), kinked tubing, empty syringe, or any restriction to flow
- If no occlusion is present, press CONFIRM
- Press START/STOP button to resume infusion
- Verify fluid is flowing in drip chamber
- A checkmark appears on standard display to indicate CONFIRM has been pressed