CCP QuickRef Handbook
[EMERG]
[ABNORM]
Home
MedSystem III Pump
Alarm: Air In Line
Meaning:
- Air detected in fluid pathway during infusion
- OR air sensor is dirty
Response:
- Verify tubing collar is fully seated in air sensor recess
- Verify tubing in air sensor recess is not damaged, twisted or dirty
- Press ClrAir softkey on Alarm Info page. Three beeps means bubble is past sensor
- If air still present: remove cassette and manually clear air
- If no air present: clean air sensor
- Press START/STOP button to resume infusion