MedSystem III Pump
Advisory: Check Air Sensor
Meaning:
- At installation of cassette:
- Air is detected in tubing
- Tubing collar is not properly seated
- Air sensor is dirty or damaged
Response:
- Verify tubing collar is fully seated in air sensor recess
- Verify tubing in air sensor recess is not damaged, twisted or dirty
- Press ClrAir on channel's Alarm Info page. Three beeps indicate air bubble is no longer in front of air sensor
- If air is still present, remove cassette and manually clear air according to hospital policy
- If no air is present, clean air sensor recess
- If pumping latch jaw is visibly broken, contact qualified service personnel